100% of user TRUEyou forearm values fell within 0.83 mmol/L of the YSI results at
glucose levels < 5.55 mmol/L and within 15% at glucose levels > 5.55 mmol/L.
Forearm Samples (user vs. YSI) for glucose concentrations <5.55 mmol/L
0.28 mmol/L 0.56 mmol/L 0.83 mmol/L
16/28 (57%)
23/28 (82%)
28/28 (100%)
Forearm Samples (user vs. YSI) for glucose concentrations >5.55 mmol/L
5%
10%
15%
42/72 (58%)
61/72 (85%)
72/72 (100%)
Forearm Samples for glucose concentrations between 1.1-33.3 mmol/L
0.83 mmol/L or + 15%
100/100 (100%)
Parkes Error Grid:
100% of individual forearm glucose measured values
performed by the users fell within Zone A of the Parkes Error Grid (PEG).
User Performance Evaluation:
A study evaluating glucose values from
fingertip capillary blood samples obtained by 100 lay persons showed the
following results: 100% 0.83 mmol/L of the medical laboratory values
at glucose concentrations below 5.55 mmol/L and 99% 15% of the
medical laboratory values at glucose concentrations at or above 5.55 mmol/L.