Instruction Manual MA-3000+RD-5+SC-5
NIC-TD-0000015-05
46
6. Option specification
6.1. RD-5
6.1.1. Performance
Number of reagent pumps
3
Type of reagent pump
Tube pump
Diameters of pump tube
3 mm (ID)
5 mm (OD)
Maximum flow rate
120 mL/min. (0.28 mL per revolution)
Dispensing quantity
Variable control by MA-3000
Driving power source
Supplied as driving signals from MA-3000
6.1.2. Specification of external connection
Connector
D-sub 9-pin plug
Connection cable
D-sub 9-pin socket, straight connection
6.1.3. Dimensions and weight
Dimensions
180 (W)
230 (D)
80 (H) (excluding protrusions)
Weight
Approximately 2 kg
6.1.4. Power supply
Power supply other than the driving signal is unnecessary.
6.1.5. Standard accessories
Item
Description
Qty
Knurled screw
For fixing SC-5 (M3
8)
1
Reagent bottle
110 mL screwed bottle with tube
3
Reagent bottle stand
Stand for 3 reagent bottles
1
Set of tubes
For in/out connection of reagent pumps
1
Pump tube
Spare part for pump tube
3
Communication cable
Dedicated communication cable for connection to
MA-3000
1
Summary of Contents for MA-3000 RD-5+SC-5