5.4. Replacement of components
5.4.1. Bubbler
1) Referring to "3Fixing of cap unit", remove the screws fixing the cover and
remove the arm covers F and R.
2) Hold the PharMed tube for the bubbler tube, and pull it up with the bubbler.
3) Pull the PharMed tube out of the bubbler.
4) Pull the bubbler out of the cap unit.
5) Insert the new bubbler in the cap unit, and make the bubbler inserted in the
PharMed tube for the bubbler tube (insertion depth: 10 mm).
6) As shown in the figure, adjust the position of the bubbler so that it is set at
108+/-1 mm from the tip of the cap.
7) Attach the arm covers F and R as before.
5.4.2. Seal packing
Pull out the packing downward and attach a new one. The packing has a smooth
surface and a sponge surface. Attach it with the smooth surface placed downward.
PharMed tube
Cap unit
Bubbler
Seal Packing
(Place the smooth
surface downward.)
Bubbler tube