P-400R Digital Microscope Instructions
89
9
T
roubleshoot
ing
Image observation/capturing
Message
Meaning
Action
Ref.
P OVER
Overexposure is
detected in Program
AE.
Since the object is too bright, an
appropriate exposure cannot be
obtained in Program AE. Adjust the
illumination.
28
P UNDER
Underexposure is
detected in Program
AE.
Since the object is too dark, an
appropriate exposure cannot be
obtained in Program AE. Adjust the
illumination.
28
S OVER
Overexposure is
detected in
Shutter-priority AE.
Since the object is too bright or the
shutter speed is too slow, an
appropriate exposure cannot be
obtained. Adjust the illumination or
set the shutter speed to fast.
28
37
S UNDER
Underexposure is
detected in
Shutter-priority AE.
Since the object is too dark or the
shutter speed is too fast, an
appropriate exposure cannot be
obtained. Adjust the illumination or
set the shutter speed to slow.
28
37
Image cannot
be saved.
An error occurred
during image recording
and the image could not
be stored.
Access to the SD card is possible,
but image recording has failed.
Replace the SD card to another one
or initialize the SD card and try
again.
12
77
White Balance
failed.
White balance could
not be obtained.
Prepare a white object, make
adjustment so that the object fills
within the metering area, and try
again.
40
Do you stop to
image at
interval?
Button operation was
performed during
interval shooting.
To stop interval shooting, select
[Yes], and to continue, select [No].
42
63
A folder already
exists.
When creating a folder,
you specified a name
which has already been
used for an existing
folder.
Create a folder with another name.
51
Select active
folder.
The folder set as an
active folder which was
used the last time
disappeared.
Select an active folder or create a
new folder.
51