75
Maintenance and Storage
5
Chapter 5
Maintenance and S
torage
1
Replacing the Lamp (for Ci-S)
CAUTION
•
Beware of burns: Wait until the lamp and nearby parts have cooled before replacing the lamp.
•
Beware of electrical shock: Turn off the power switch and unplug the power cord from the wall outlet.
•
Beware of abnormal heat generation: Use only the designated lamp.
•
Beware of soiling: Avoid touching the glass surface of the lamp with bare hands. Soiling may reduce the service life
of the lamp.
•
Lamphouse cover: Make sure the lamphouse cover is securely fitted to the lamphouse after lamp replacement.
•
Used lamps: Do not break used lamps. It should be disposed of as industrial waste, according to local regulations
and rules.
(1) Remove the lamphouse cover on the back of the
microscope. Remove the old lamp.
(2) Attach a new lamp. Avoid touching the glass surface of the
lamp with your bare hands. Use only the specified lamp
(PHILIPS 5761).
(3) Restore the cover back to its original position. Engage the
cover hook in the slot on the rear of the unit in the opposite
direction indicated by the arrow in the figure.
DSC
Replacing the lamp
Lamp
Lamphouse cover
Cover hook
Slot