Display
Display
Display
Display (
(
(
( DL
DL
DL
DL -
-
-
- D5
D5
D5
D5 -
-
-
- DISP
DISP
DISP
DISP)
)
)
)
① Display Panel
Two line display shows angle measurements, battery status, and
communication connection status.
Angle
DL-S3 measured value is displayed in mm/m.
Top row shows data for DL-S3 connected to A channel;
bottom row shows data for DL-S3 connected to B channel.
If DL-S3 units are set to "DEG", the display will indicate "Mismatch".
Please make sure the DL-S3 is set to mm/m.
If the DL-S3 measurement is out of range, it will output an error, and
an error will be displayed. For positive out of ranger error, "+Error"
is displayed, for negative, "-Error" is displayed.
(Refer to the DL-S3 Manual for measurement range and errors.)
The DL-S3 may send out a valid measurement which appears outside the
specified range due to the function of the 0-Cal and 1/2-Cal operations.
(Ex:If 0-Cal operation is performed when the DL-S3 is at angle of
-4mm/m, it will set the display to 0. Then when the DL-S3 is tilted
to +4mm/m, the display will read +8mm/m.)
Battery Low
When battery voltage begins to drop, the display will alternate between
measurement and "Low Power Battery" indication.
Replace battery immediately.
Establishing Communication
When communication with the Transmitter is not established, (due to
Transmitter power OFF or out of range) the display will show
"
‐‐‐‐
"(16 char, 2 lines.)After about 5 min., display will turn off.
② Power ON button
When first pressed, "ON-SW keeping" will display; continue to hold button.
When power is ON, display will change to "ON keep OK"
③ Power OFF button
When pressed, "OFF-SW keeping" will display.
Press and hold (approx. 2 seconds) to turn power OFF.
If there is no communication with the Transmitter, after about 5 min. the
Display power will automatically turn OFF.
④ Backlight button
When the LIGHT button is pressed, the display backlight will go on for about 1 min.
If the LIGHT button is pressed while the light is on, the backlight will turn off.
⑤ 0-Cal button(A channel)
Press the 0-Cal button to set the displayed reading to 0.
Press and hold for about 2 sec. to insure activation.
(Button does not work during error display.)
⑥ 1/2-Cal button(A channel)
Press the 1/2-Cal button to divide the displayed reading by 2.
Value is changed when the 1/2-Cal button is released.
Press and hold for about 2 sec. to insure activation.
(Button does not work during error display.)
⑦ 0-Cal button (B channel)
⑧ 1/2-Cal button (B channel)
Key ring
①
③
④
②
⑤
⑦
⑥
⑧
【Silicone Covers attached】