196
8.13.2.2 Lens measurement axis adjustment
1 . Display the menu screen
2 . Press [LMU adjustment].
3 . Select the Adj. tab to display the LMU
adjustment parameters.
4 . Press the LMU axis constant [deg] field.
5 . Adjust the lens measurement axis by the
procedure below.
1 ) If the trace rises up to the right,
increase the value for the LMU axis
constant [deg] parameter.
2 ) If the trace drops down to the right,
decrease the value for the LMU axis
constant [deg] parameter.
/
Trace
Pliable
cup
/
Trace
Pliable
cup
Summary of Contents for SE-9090 Supra
Page 1: ...SYSTEM EDGER SE 9090 Supra SERVICE MANUAL September 2012 Pages in total 350 LLE7D RDA002B...
Page 12: ...LLE7D RDA002B X This page is intentionally left blank...
Page 14: ...LLE7D RDA002B 2 This page is intentionally left blank...
Page 22: ...LLE7D RDA002B 10 This page is intentionally left blank...
Page 24: ...LLE7D RDA002B 12 This page is intentionally left blank...
Page 104: ...LLE7D RDA002B 92 This page is intentionally left blank...
Page 310: ...LLE7D RDA002B 298 9 3 Wiring Diagram...
Page 311: ...LLE7D RDA002B 299 9 4 Connector Cable...
Page 312: ...LLE7D RDA002B 300...
Page 313: ...LLE7D RDA002B 301 9 5 Layout Drawing 9 5 1 ASSYs 40274 2000...
Page 314: ...LLE7D RDA002B 302 40274 3000 40274 5000 40274 6000...
Page 315: ...LLE7D RDA002B 303...
Page 316: ...LLE7D RDA002B 304 40274 4000 40274 8000...
Page 317: ...LLE7D RDA002B 305 40274 7100 40274 7200 40273 7200 2 2 5 2 5 5 2 2 2 5...
Page 318: ...LLE7D RDA002B 306 9 5 2 Boards S N 700019 and earlier...
Page 319: ...LLE7D RDA002B 307 S N 700020 and later...
Page 320: ...LLE7D RDA002B 308 9 5 3 Cable ASSYs...
Page 321: ...LLE7D RDA002B 309 9 5 4 Other electrical components...
Page 322: ...LLE7D RDA002B 310...
Page 350: ...LLE7D RDA002B 338 This page is intentionally left blank...