TEST METHOD
Negative Relative Accommodation (NRA) Test
4.11 Negative Relative Accommodation (NRA) Test
Use: To measure the negative relative accommodation when both eyes are converged at a specified
working distance
Chart: Near letter chart with a horizontal line isolation at 40 cm away from the patient’s eye (near point
chart 7)
Confirm that the “SPH Far
Near” parameter is set to SPH + ADD.
For the Plus Package, see "5.2.6 #21 Negative relative accommodation (NRA)" (page 218).
1
Press
while pressing
.
[NPC], [NPA], [NRA], and [PRA] appear.
When the NPA test has been completed just before, skip this step.
2
Press [NRA].
Both measuring windows are opened.
*29
The NRA window is displayed.
3
Gradually turn the dial counterclockwise
and press [Blur] when the chart appears
blur.
The [Blur] indication is highlighted. It
means that the value has been stored.
4
Gradually turn the dial clockwise and press
[Recov] when the chart appears clearly again.
The [Recov] indication is highlighted. It means that the value has been stored.
Pressing [Blur] or [Recov] again clears the stored data.
*29.
Occluding either measuring window with
or
measures the near point of accommodation for either eye.
161221687130 承認済
Uncontrolled copy when printed.
Summary of Contents for RT-5100
Page 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Page 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Page 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Page 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Page 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...