TEST METHOD
Near Point of Accommodation (NPA) Test
4
4.10 Near Point of Accommodation (NPA) Test
Use: To measure the near point of accommodation
Chart: The highest VA letter chart that the patient can read correctly at the position of 40 cm away
from the near point chart
Confirm that the “SPH Far
Near” parameter is set to SPH + ADD.
For the Plus Package, see "5.2.13 #19 Near point of accommodation (NPA)" (page 225).
1
Press
while pressing
.
[NPC], [NPA], [NRA], and [PRA] appear.
When the NPC test has been completed just before, skip this step.
2
Press [NPA].
Both measuring windows are opened.
*28
3
Gradually bring the chart closer to the
patient’s eye and measure the distance
from the chart to the root of the nose at the
point where the chart appears blur.
4
Enter the distance with the dial.
Also possible by pressing
or
.
The distance from 1 to 100 cm can be
entered in 1 cm increments.
The accommodation is calculated automatically and the result is printed with the other data.
For example: Enter 33 cm as the distance when ADD = +2.0 D (addition powers are entered
in far mode)
Accommodation power = 1/0.33 – (+2.0) = 1.03 D
NPA = 33 cm: 1.03 D
*28.
Occluding either measuring window with
or
measures the near point of accommodation for either eye.
161221687130 承認済
Uncontrolled copy when printed.
Summary of Contents for RT-5100
Page 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Page 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Page 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Page 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Page 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...