![Nidek Medical LT-980 Service Manual Download Page 21](http://html1.mh-extra.com/html/nidek-medical/lt-980/lt-980_service-manual_1688033021.webp)
21
5.1.3.3 ERROR19: Trace cancel
Conduct tracing again.
5.1.3.4 ERROR20: H-axis encoder error
Perform the procedure as in “7.1.1 Tracer ASSY (40601-0010)” (p33).
5.1.3.5 ERROR21: No frame
5.1.3.6 ERROR22: No pattern
Set a pattern and conduct tracing again.
5.1.3.7 ERROR23: Frame moved during tracing
Conduct tracing again while holding the frame so that it does not move.
The tracer ASSY (40601-0010) is defective.
When repairing the tracer ASSY (40601-0010), refer to the service manual for the tracer
ASSY (LETR-1).
Is tracing conducted with no frame set?
Yes
Set a frame and conduct tracing again.
No
Is the set frame smaller than the minimum
traceable range?
Yes
Tracing is not possible.
No
Conduct tracing again.
Summary of Contents for LT-980
Page 1: ...Satellite Tracer LT 980 SERVICE MANUAL October 2012 Pages in total 82 LLT9A RDA002A...
Page 2: ......
Page 6: ...LLT9A RDA002A 6 This page is intentionally left blank...
Page 8: ...LLT9A RDA002A 8 This page is intentionally left blank...
Page 12: ...LLT9A RDA002A 12 This page is intentionally left blank...
Page 14: ...LLT9A RDA002A 14 This page is intentionally left blank...
Page 32: ...LLT9A RDA002A 32 This page is intentionally left blank...
Page 44: ...LLT9A RDA002A 44 This page is intentionally left blank...
Page 57: ...LLT9A RDA002A 57 5 Remove the standard frame 40601 7500 40601 7500 Circumference...
Page 69: ...LLT9A RDA002A 69 9 3 Wiring Diagram...
Page 70: ...LLT9A RDA002A 70 9 4 Connector Cable 1 44801 CA01 2 44801 CA02 3 44801 CA03 4 44801 CA04...
Page 80: ...LLT9A RDA002A 80 This page is intentionally left blank...
Page 81: ......