67
CHECK AND ADJUSTMENT PROCEDURES: Groove Depth Check and Adjustment
3.13.2 Groove depth adjustment
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Grinding” tab.
3
Press Tab 3.
4
Press the numeric field of “Groove depth”.
The numeric keypad appears.
5
Enter a value with the numeric keypad and
change the parameter value.