40
CHECK AND ADJUSTMENT PROCEDURES: Bevel Position Check and Adjustment
3.7.2
Bevel position adjustment
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Grinding” tab.
3
Press Tab 3.
4
Press the numeric field of “Bevel position
adjustment for auto mode”.
The numeric keypad appears.
5
Enter a value with the numeric keypad and
change the parameter value.