
91
PARAMETER SETTINGS AND MAINTENANCE: Parameter Settings
●
SAGITTAL
SAGITTAL parameter screen
Parameter option
Setting contents
21. SAGITTAL
(Sagittal measurement)
YES / NO
Sets whether to conduct sagittal measurement after KM measurement.
When “YES” is selected, the SAG icon is displayed on the AR/KM measure-
ment screen.
22. SAGIT AXIS
(Sagittal cylinder axis)
AXIS / FIX
Sets whether to convert the sagittal measurement for the axis of the steep-
est meridian obtained by KM measurement.
•
AXIS: Converted to axis value of KM measurement
•
FIX: Measured values are used
23. SAGIT PRINT
(Printing of sagittal data)
ALL / SHORT
Selects the format of sagittal measured data to be printed.
•
ALL: All data are printed
•
SHORT: Only sagittal curvature values and the total eccentricity are
printed.