8.8
KM CALIBRATION
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
8.8.1 KM CALIB
1 . Attach the KM calibration jig (30670-0300)
2 . Display the CALIB MODE SELECT screen (for KM), and select KM CALIB
* When the software version is earlier than Ver. 2.03, press the reset button.
* When the software version is Ver. 2.03 or later, the reset button is not displayed.
3 . Perform KM CALIBRATION as follows:
1 ) Align the relevant steel ball of the
KM calibration jig (30670-0300) in
the order of the following numeric
values so that the focusing indicator
displays the optimum condition.
a . 7.94 mm
b . 5.95 mm
c . 9.13 mm
2 ) Press the start button to perform KM
CALIBRATION.
* Perform calibration of all three
steel balls.
3 ) Press the SAVE button to save the setting.
4 . Press the exit
button.
-/%#.+$4#6+10
OO
%#.+$4#6+10(+0+5*
Summary of Contents for AR-310A
Page 2: ......
Page 10: ...XAR17ARDA001G E 10 Return to previous page This page is intentionally left blank ...
Page 62: ...XAR17ARDA001G E 62 Return to previous page This page is intentionally left blank ...
Page 68: ...XAR17ARDA001G E 68 Return to previous page This page is intentionally left blank ...
Page 102: ...XAR17ARDA001G E 102 Return to previous page This page is intentionally left blank ...
Page 186: ...XAR17ARDA001G E 186 Return to previous page 9 3 1 2 30601 E000 AR 310A and ARK 510A ...
Page 190: ...XAR17ARDA001G E 190 Return to previous page 30601 EA14 ...
Page 192: ...XAR17ARDA001G E 192 Return to previous page 30601 EA23 30601 EA24 ...
Page 194: ...XAR17ARDA001G E 194 Return to previous page 30601 EA30 30601 EA31 30601 EA32 ...
Page 196: ...XAR17ARDA001G E 196 Return to previous page 30601 EA51 30601 EA52 ...
Page 217: ......