Characteristics

 describe values that are relevant to the use of the model under stated operating

conditions but are not covered by the model warranty.

Typical

 specifications describe the performance met by a majority of models.

Nominal

 specifications describe an attribute that is based on design, conformance testing,

or supplemental testing.

Specifications are 

Warranted

 unless otherwise noted.

Voltage, Current, and Resistance

Notice

 

Refer to the specifications of the module with which you are using the

TB-414X to note the maximum voltage and current ratings. The specifications listed
below are the maximum for the TB-414X only.

Channel operating voltage

1

60 V DC

HI/LO pin current

2

3 A

Note

 

It is critical to ensure that the LO terminal for each channel in use is

connected when the total current from all channels exceeds 3 A. The TB-414X is
rated for the force pins carrying a maximum of 3 A average current, however,
individual pin currents above 4 A may damage or degrade the lifespan of the pins.
Systems that have a common path between the LO of each channel at the DUT end
of the cable must ensure that LO path wiring resistances are matched within ±20% to
prevent the LO terminal of one channel from carrying a significant portion of
another channel’s LO current and exceeding the 4 A individual pin limit.

Force path resistance

3

30 mΩ, typical

Leakage resistance

4

23 °C

>100 TΩ, typical

1

Maximum voltage measured between the HI, HI Sense, LO, LO Sense, or Guard terminals. SMU
specifications take precedence if the SMU model is rated for lower voltage.

2

SMU specifications take precedence if the SMU model is rated for lower current.

3

Combined resistance in the HI and LO force traces and connectors for a channel of the TB-414X.
Does not include any output resistance contributions of the SMU module or user wiring. Refer to
SMU module specifications to determine if this resistance is included in the voltage load regulation
specification or if it should be accounted for separately.

4

Applies between HI and LO terminals. Leakage performance may be degraded for operation above
70% relative humidity. When transitioning a device from a storage or operation environment with
operating humidity above 70%, allow the device to stabilize in the lower humidity environment for
several hours before use.

8

  | 

ni.com

  | 

TB-414X User Guide

Summary of Contents for TB-414X

Page 1: ...nstalling and configuring your system Caution Observe all instructions and cautions in the user documentation Using the product in a manner not specified can damage the product and compromise the buil...

Page 2: ...when you see this icon printed on the model Cautionary statements are localized into French for compliance with Canadian requirements Getting Started with the TB 414X Kit Contents and Other Equipment...

Page 3: ...ze negative effects from wiring insulation materials and contaminants The SHDB25F DB25F LL cable accessory has been designed with these considerations in mind and utilizes the TB 414X allowing for one...

Page 4: ...the stripped end of the wires into the appropriate terminals and tighten the screws of the screw terminals to a torque of 0 5 Nm 4 lb in Refer to the following pinout signal information and PXIe 414x...

Page 5: ...H 1 C H 2 Table 1 TB 414X Signals and DB25 Pin Map Screw Terminal DB25 Connector Pin LO Connector CH 0 LO 16 CH 0 LO S 3 CH 1 LO 19 CH 1 LO S 6 CH 2 LO 22 CH 2 LO S 9 CH 3 LO 25 CH 3 LO S 12 CH S GND...

Page 6: ...he strain relief upward If you are connecting only a few wires assemble the strain relief downward for greater retention Figure 3 TB 414X Strain Relief 1 2 1 Strain Relief Up 2 Strain Relief Down b In...

Page 7: ...ations These specifications apply to the TB 414X for use with the PXIe 414x The PXIe 414x is a 4 Channel SMU Refer to the module specifications at ni com manuals for more information Definitions Warra...

Page 8: ...nel at the DUT end of the cable must ensure that LO path wiring resistances are matched within 20 to prevent the LO terminal of one channel from carrying a significant portion of another channel s LO...

Page 9: ...ements of voltages from specially protected secondary circuits Such voltage measurements include signal levels special equipment limited energy parts of equipment circuits powered by regulated low vol...

Page 10: ...ronmental Guidelines Notice This model is intended for use in indoor applications only Environmental Characteristics Temperature and Humidity Temperature Operating 0 C to 55 C6 Storage 40 C to 71 C 6...

Page 11: ...NI customers For additional environmental information refer to the Commitment to the Environment web page at ni com environment This page contains the environmental regulations and directives with whi...

Page 12: ...RMA by NI does not constitute export authorization The user must comply with all applicable export laws prior to exporting or re exporting this product See ni com legal export compliance for more inf...

Page 13: ...f help such as tutorials and examples Visit ni com services to learn about NI service offerings such as calibration options repair and replacement Visit ni com register to register your NI product Pro...

Page 14: ...LAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how...

Reviews: