6.
Restart your computer.
7.
Verify that the device appears in MAX.
What Should I Do if the Module Fails the Self-Test?
1.
Restart the system.
2.
Launch MAX, and perform the self-test again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test again.
What Should I Do if the Instrument Does Not Initialize?
Failure to initialize may indicate a problem with module interconnections or with MAX. If the
niRFSA Initialize VI or niRFSG Initialize VI returns an error and the PXIe-5831 fails to
initialize, complete the following steps:
1.
Reconnect the PXIe-5831 front panel cables securely.
2.
Power on your system and run the MAX configuration and self-test procedures.
PXIe-5831 Getting Started Guide
|
© National Instruments
|