Appendix A
Specifications
A-2
ni.com
NI 6052E Accuracy Information
Input coupling.........................................DC
Max working voltage
( common-mode)........................Each input should remain
within ±11 V of ground
Overvoltage protection ...........................±25 V powered on,
±15 V powered off
Nominal Range (V)
Absolute Accuracy
Relative
Accuracy
% of Reading
Offset
(
µ
V)
Noise +
Quantization (
µ
V)
Temp
Drift
(%/
°
C)
Resolution (
µ
V)
Positive
FS
Negative
FS
24 Hours
90 Days
1 Year
Single
Pt.
Avg.
Single
Pt.
Avg.
10
–10
0.0304
0.0312
0.0321
1067
981
87.0
0.0006
1145
115
5
–5
0.0054
0.0062
0.0071
536
491
43.5
0.0001
573
57.3
2.5
–2.5
0.0304
0.0312
0.0321
271
245
21.7
0.0006
286
28.6
1
–1
0.0304
0.0312
0.0321
111
98.1
8.7
0.0006
115
11.5
0.5
–0.5
0.0304
0.0312
0.0321
58.1
56.2
5.0
0.0006
66.3
6.6
0.25
–0.25
0.0304
0.0312
0.0321
31.6
32.8
3.0
0.0006
39.2
3.9
0.1
–0.1
0.0304
0.0312
0.0321
15.6
22.4
2.1
0.0006
27.7
2.8
0.05
–0.05
0.0304
0.0312
0.0321
10.3
19.9
1.9
0.0006
25.3
2.5
10
0
0.0054
0.0062
0.0071
536
491
43.5
0.0001
573
57.3
5
0
0.0304
0.0312
0.0321
271
245
21.7
0.0006
286
28.6
2
0
0.0304
0.0312
0.0321
111
98.1
8.7
0.0006
115
11.5
1
0
0.0304
0.0312
0.0321
58.1
56.2
5.0
0.0006
66.3
6.6
0.5
0
0.0304
0.0312
0.0321
31.6
39.8
3.0
0.0006
48.2
3.9
0.2
0
0.0304
0.0312
0.0321
15.6
22.4
2.1
0.0006
27.7
2.8
0.1
0
0.0304
0.0312
0.0321
10.3
19.9
1.9
0.0006
25.3
2.5
Note:
Accuracies are valid for measurements following an internal E Series calibration. Averaged numbers assume dithering
and averaging of 100 single-channel readings.
Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C
of external or factory calibration temperature.