Table 5.
Values for Verifying the Offset of the Main Analog Path
Iteration
Gain
Ideal
Positive
Offset
(Volts)
Ideal
Negative
Offset
(Volts)
As-Found
Limits (Volts)
As-Left
Limits (Volts)
1
12.000000
6.000000
- 6.000000
±0.051500
±0.021500
2
10.000000
5.000000
-5.000000
±0.043000
±0.018000
3
7.000000
3.500000
-3.500000
±0.030250
±0.012750
4
5.000000
2.500000
-2.500000
±0.021750
±0.009250
5
3.500000
1.750000
-1.750000
±0.015375
±0.006625
6
2.500000
1.250000
-1.250000
±0.011125
±0.004875
7
2.000000
1.000000
-1.000000
±0.009000
±0.004000
8
1.650000
0.825000
- 0.825000
±0.007513
±0.003388
9
1.250000
0.625000
- 0.625000
±0.005813
±0.002688
10
0.850000
0.425000
- 0.425000
±0.004113
±0.001988
11
0.600000
0.300000
- 0.300000
±0.003050
±0.001550
12
0.415000
0.207500
- 0.207500
±0.002264
±0.001226
13
0.300000
0.150000
- 0.150000
±0.001775
±0.001025
14
0.205000
0.102500
- 0.102500
±0.001371
±0.000859
15
0.150000
0.075000
- 0.075000
±0.001138
±0.000763
16
0.105000
0.052500
- 0.052500
±0.000946
±0.000684
17
0.075000
0.037500
- 0.037500
±0.000819
±0.000631
18
0.055000
0.027500
- 0.027500
±0.000734
±0.000596
19
0.037500
0.018750
- 0.018750
±0.000659
±0.000566
20
0.026000
0.013000
- 0.013000
±0.000611
±0.000546
21
0.018500
0.009250
- 0.009250
±0.000579
±0.000532
22
0.013000
0.006500
- 0.006500
±0.000555
±0.000523
23
0.009000
0.004500
- 0.004500
±0.000538
±0.000516
24
0.006500
0.003250
- 0.003250
±0.000528
±0.000511
Note
:
Error Positive Offset Value
= (
Measured Positive Offset Value
) - (
Ideal Positive Offset Value
)
Error Negative Offset Value
= (
Measured Negative Offset Value
) - (
Ideal Negative Offset Value
)