KTMED OXYPALM SpO2 monitor 

Supplemental Operators Information 

 

 
 
 

 
OXYPALM is indicated for use in measuring and displaying functional oxygen saturation and pulse 
rate for neonatal, paediatric and adult use. 

The device should be used for spot-checking only and 

isn’t suitable for unattended continuous monitoring, due to limitations of the Audio Alarm. 

 

Functions & Display: 

 

Power button, Press (>2 Seconds) to turn device ON.  All LEDs are shown, followed by device 
software version.  Press Power button (>2 Seconds) to turn device OFF. 

 

Upper numeric display = Patient SpO2 reading. 
 
Lower numeric display = Patient Pulse rate. 
 
Green Segment bar = Pulse Strength Indicator.  
 
“SEn” “OPn” = “Sensor Open”, Sensor not connected to device. 
 
“Fin” “OPn” = “Finger Open”, Sensor is not attached to the patient. 
 
“POA” “OFF” = “Power OFF”, Device is about to turn OFF. 
 
“0” “0” = If both SpO2 and Pulse rate display numeric are Zero then device is 

Searching

 for a 

signal. 
 
Numeric Flashing = If both SpO2 and Pulse rate display numeric are flashing then this 
indicates a 

Noisy Signal

, normally due to patient movement or poor positioning of the sensor. 

 
Green LED (solid) = Audio Alarm is 

Silenced

To activate 

Alarm Silence, 

Press M button (short press).  Press M button to re-activate audio. 

Note: Audio Alarm re-activates after the set Alarm Interval time is reached: OFF, 30sec, 60Sec, 
90Sec, 120Sec. Warning: If Alarm Interval time is set to OFF device remains permanently silenced. 
 
Red LED Flashing = 

Alarm condition

 
Green LED (solid) + Red LED Flashing = 

Silenced Alarm condition

.  Visual Alarm condition only. 

 
Battery Low LED = Batteries require replacement. 