2.2 Installation
FRA5014
2-3
2.2 Installation
2.2.1 General precautions for installation
Note the following precautions to prevent damage to the
FRA5014
.
•
The
FRA5014
uses a fan for forced-air cooling. Air intake occurs via the bottom front
panel of the unit and air exhaust occurs via the central rear panel. Do not obstruct
either the air intake or exhaust areas.
•
If the
FRA5014
is used with its rear panel facing downward (installed vertically), it
may topple easily. Always install it horizontally (with its bottom facing the floor).
◼
Handling of panels and case
The
FRA5014
’s front panel is made of plastic. Be careful to prevent damage that may be
caused by sharp objects or high temperatures.
When the panel or case surfaces become dirty, use a soft cloth to wipe them. For stubborn dirt,
dip the cloth in a neutral detergent and wring it out well before wiping off the dirt . Avoid use
of organic solvents such as paint thinner or benzene, or pre-moistened wipes, as they may cause
discoloration, clouding, and/or peeling of the surface coating.
2.2.2 Installation conditions
Install the
FRA5014
in a location that meets the following conditions for temperature and humidity
ranges.
Performance guarantee
+5 to +35
C, 5 to 85 %RH (Absolute humidity of 1 to 25 g/m
3
, with
no condensation)
Operation
0 to +40
C, 5 to 85 %RH (Absolute humidity of 1 to 25 g/m
3
, with no
condensation)
Storage
−
10 to +50
C, 5 to 95 %RH (Absolute humidity of 1 to 29 g/m
3
, with
no condensation)
Do not install the FRA5014 in the following locations:
•
Location exposed to direct sunlight or near a heat source
•
Locations where dust, salt, and/or metallic powder is prevalent
•
Locations where corrosive gas, steam, and/or oily smoke is prevalent
•
Location exposed to excessive vibration
•
Within or close to sources of strong magnetic fields
•
Close to sources of pulsing noise.
When measurement precision is important, warm up the
FRA5014
at least two hours before using
it, then run self calibration before starting measurements. If the ambient temperature has
changed more than 5 °C, or if the
FRA5014
has been used continuously for more than 24 hours,
running self calibration again is recommended.
Keep the mains supply cord separated from the signal cable. If the mains supply cord and signal
cable get too close, operation faults or noise may lead to inaccurate measurements.
!
CAUTION
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......