4.2 Using Sequence Function
DP060LM/DP120LM/DP180LM
73
Step Behavior (Const, Keep, Sweep)
For the intra-Step parameter other than waveforms and phases, the following three patterns can be set
as the method to change values in the step.
Constant (Const)
Fixed to the value set in the step.
Keep
Keeping the value immediately before transitioning to the step. That is, the value is changed
depending on the status of the previous step. Although the parameter value can be set even if the
operation type setting is Keep, the parameter setting value is ignored when sequence is
executed.
Sweep
It changes linearly from the value immediately before transitioning to the value set in the step,
taking the Step Times. In other words, it reaches the setting value after the Step Time passes in
the step. The starting value is changed depending on the status of the previous step.
Previous step
Next step
Previous step
Next step
Previous step
Next step
Used as the setting
value in the next step
Keeps the final value
of the previous step
Changes from the final value of
the previous step to the setting
value of the next step
Cons
t
Keep
Sweep
Setting
value
Setting
value
Figure 4-2 Step Behavior
--------
Notes
------------------------------------------------------------------------------------------------------------
The Step Behavior for the waveform and phase is always Constant (Const).
Keeps as many intra-Step parameter setting values as the number of phases in each step based
on the output phase mode.
Cannot be set using line voltage in the single-phase three-wire output and three-phase output.
The rate of crest factor/clip ratio cannot be changed though a clipped sine wave can be selected
as an output waveform in Sequence function. Set these parameters beforehand using
Continuous function.
Sequence function displays the measured value in Sequence function, which is equivalent to
that of the Continuous function. However, harmonic current measurement (see 4.4) cannot be
displayed.
-----------------------------------------------------------------------------------------------------------------------------
Step Behavior
Summary of Contents for DP060LM
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Page 4: ......
Page 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Page 24: ......
Page 224: ......
Page 244: ......
Page 264: ......
Page 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...