11.20 Simulation
DP060LM/DP120LM/DP180LM
267
11.20 Simulation
This allows you to simulate power line abnormalities, such as blackout, voltage rise, voltage drop,
abrupt phase change, and abrupt frequency change. It can be used only for AC, sine wave, and
ACDC-INT. For the polyphase output only the balanced mode is available. Note that it does not
support IEC or other standards test. When performing the tests specified by the standards, use
appropriate peripheral equipment.
All
models
Number of memories
5 (non-volatile)
Number of steps
6 (Initial, Normal 1, Trans 1, Abnormal, Trans 2, Normal 2)
Setting range of Step
Time
0.0010 s to 999.9999 s (0 s is available only for the Transition Step)
Parameter Output
range
(The above item is common within the Simulation)
AC voltage
Frequency
Waveform (sine wave only)
Start Phase (excluding the Transition Step)
Stop Phase (excluding the Transition Step)
Synchronous step output (2bit)
Trigger output
Repeat count (1 to 9999 times or infinite)
初期
定常1
移行1
異常
移行2
定常2
初期
トリガ設定例
ステップ同期1出力例
ステップ同期2出力例
Initial
Initial
Normal1
Normal2
Trans1
Trans2
Abnormal
Trigger setting example
Step sync 1 output example
Step sync 2 output example
Simulation control
Start
Stop
Summary of Contents for DP060LM
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Page 4: ......
Page 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Page 24: ......
Page 224: ......
Page 244: ......
Page 264: ......
Page 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...