8.2 Using in Combination with DIP or RIN
DP060LM/DP120LM/DP180LM
223
8.2
Using in Combination with DIP or RIN
When using this product in combination with DIP or RIN, connect between CONTROL I/O
connector and CONTROL IN connector of this product with the attached control cable (25 pin) and
connect CONTROL SIGNAL connector of this product and DIP or RIN with the control cable
included in DIP or RIN. Also set external control function to DevCtrl.
For detail on how to use DIP or RIN, refer to the respective instruction manual.
8.2.1
The Case the Voltage Drop Due to Wiring is Small
(a)
The case measuring with the output terminal of this product
The feedback cable included in DIP or RIN and the FEEDBACK connector of DIP or RIN are
not used. Turn off the remote sensing function of this product.
(b)
The case measuring with the sensing input terminal of this product
This product measures the voltage at the FEEDBACK connector of DIP or RIN. Connect
between the FEEDBACK connector of DIP or RIN and the sensing input terminal of this
product with the feedback cable included in DIP or RIN. Turn on the remote sensing function of
this product.
Turn off this product including its peripherals before connecting or disconnecting the cable. Be
sure to turn off this product and switch off the breaker at the distribution board. Otherwise, an
electric shock might occur and it may damage the load and this product.
For the polyphase output, cables should be connected to the sensing input terminals N1, N2, and
N3 for each phase. These terminals do not have the same electric potential within this product.
8.2.2
Compensating the Voltage Drop by the FEEDBACK Connector of DIP
or RIN
(a)
Compensate when setting remote sensing function to FB
By changing the voltage feedback detection point from the inside of this product to the external
(sensing input terminal), the output voltage at the FEEDBACK connector of DIP or RIN can be
compensated.
Connect between the FEEDBACK connector of DIP or RIN and the sensing input terminal of
this product with the feedback cable included in DIP or RIN. Figure 8-1 shows the control cable
connection diagram of DP060LM and RIN (DP4162) in single-phase output and Figure 8-2
shows that of DP060LM and DIP (DP4172) in three-phase output for reference.
Set the remote sensing function of this product to FB. (See
4.11
.)
For connection diagram of the other model, refer to its manual.
WARNING
CAUTION
Summary of Contents for DP060LM
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Page 4: ......
Page 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Page 24: ......
Page 224: ......
Page 244: ......
Page 264: ......
Page 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...