4.
Advanced Operation
PROGRAMMABLE AC POWER SOURCE
108
Voltage change simulation
In this simulation, the 50 Hz 100 V power source repeats the voltage change three times at 10-second
intervals, while the voltage drops to 70 % and then linearly recovers in 0.5 seconds during one cycle
as shown in Figure 4-13. For this case, set so that the total of Normal 1 and Normal 2 Step Times
becomes 10 seconds and set the repeat count to 3. Set the Trans 1 Step Time to zero (Table 4-7).
正常時
100 V
低下時
70 V
周波数
: 50 Hz
電圧
上昇時
5
秒
0.02
秒
0.5
秒
5
秒
3
回繰り返し
At drop
Frequency: 50Hz
Repeat 3 times
5 sec
5 sec
0.5 sec
At normal
At voltage
rise
0.02 sec
Figure 4-13 Voltage Change Simulation Example
Table 4-7 Voltage Change Simulation Edit Example
Step Initial
Normal 1
Trans 1
Abnormal
Trans 2
Normal 2
Step Time
—
5 s
0 s
0.02 s
0.5 s
5 s
AC voltage
100 V
100 V
—
70 V
—
—
Frequency
50 Hz
50 Hz
—
50 Hz
—
—
Start
phase
OFF OFF — OFF — OFF
Stop
Phase
OFF OFF — OFF — OFF
Repeat count
2 times
Summary of Contents for DP060LM
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Page 4: ......
Page 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Page 24: ......
Page 224: ......
Page 244: ......
Page 264: ......
Page 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...