OPERATOR'S MANUAL
73
8. TECHNICAL DATA
8.1. ELECTRICAL CHARACTERISTICS
Input voltage
115
–
240 VAC Single phase
Power supply frequency
50 / 60 Hz
Standby power (absorption in stand-by mode)
20 W
Maximum current absorbed in operating conditions
20A @ 115V (20A Fuse)
12A @ 240V (12A Fuse)
Maximum apparent line resistance
0.5 Ω @ 240 V
-
0.25 Ω @ 115 V
Maximum peak input anode power
1184 W (e.g. 74 kV / 16 mA)
Current / voltage combination for the maximum output
power
74 kV / 16 mA (1184 W)
Column motor
25 s on
400 s off
Kinematic motor
1/20
8.2. RADIOLOGICAL CHARACTERISTICS
Generator voltage
2D: 60
–
85 kV
3D: 90kV pulsed mode
Anode current
2
–
16 mA
Type of generator
Constant potential (DC)
Generator frequency
100
–
180 kHz
Maximum continuous input anode power
42 W
Maximum deviations from declared values
kV: < 5%
mA: < 10%
ms: < 5% + 50 ms
mAs: < 10% + 0.2 mAs
Linearity error < 0.2
Coefficient of variation < 0.05
X-ray tube
CANON / TOSHIBA D-0510SB
CEI OPX/105-12
Focal spot size
0.5 mm according to IEC 60336
Anode material
Tungsten (W)
Anode inclination
10° (CANON / TOSHIBA D
-0510SB)
12° (CEI OPX/105
-12)
Anode thermal capacity
30 kJ (CEI tube)
–
35 kJ (CANON / TOSHIBA tube)
Temperature curves of cylinder block
Inherent filtration (including supplementary filtration)
2D: > 2.5 mm Al @ 85 kV
3D: 6.5 mm Al at 90 kV
Half value layer (HVL)
2D: > 2.9 mm @ 80 kV
Leakage Technique Factor (LTF)
90 kV; 0.47 mA
Stray radiation
<0.88mGy / h at 1 metre from focal spot at 90 kV @
42 W
Exposure time (PAN/CEPH/CBCT)
2D: 1 s - 18 s continuous emission (depending on the
2D examination type selected)
3D: 1 s
–
10.4 s pulsed emission