OPERATOR'S MANUAL
EN
7) Once the wanted position has been reached, touch NEXT to go to the EXECUTION OF EXAMINATION screen.
8) Go to the PC workstation.
When the patient accesses the machine during the “Patient Positioning” or the “Execution of Examination”
(and anyway before starting the examination), make sure that the X-ray device is not accidentally hit: in this
case, it is recommended to make the patient exit the machine and reposition the X-ray device, returning to
the “Examination Settings” screen and repeating the procedure.
6.5. PERFORMING THE EXAMINATION
- Visually check the correct position of the patient and make sure that the green LED is steadily on and in the X-ray
remote control.
- Tell the patient not to move during the examination, to breath slowly and regularly,
- Make all the unnecessary persons move away from the area exposed to radiation and, when required, go behind
the suitable protection.
The console EXECUTION of examination screen shows
the X-ray parameters set and the exposure progression.
The user can see the amount of dose being
administered and how much time is left before the
examination ends
Follow the instruction of the Acquisition Operations section in the NNT manual to complete the image acquisition.
Refer to “Extended View” mode f
or the acquisition with fields of view of the Extended View type.
-
Press the X-Ray remote Control to perform the
emission and keep it pressed during the whole
duration of the examination. The duration of the
examination is indicated by the yellow flashing
LED of the X-ray remote control device. The X-ray
emission is signalled with an audible signal.