22 System
Operation
4.1.2 Status
The table below shows the meaning of the various colors of SYSTEM, CH1
and CH2 status LED’s.
Item
Level
Category
Comment
Ch1
LED
Ch2
LED
System
LED
1
System
Unpowered AC not connected or Power Switch off
OFF
OFF
OFF
2
System
Unpowered Blown fuse
OFF
OFF
OFF
3
System
Unpowered Defective power supply
OFF
OFF
OFF
4
System
Powered
Controller performing self-check
-
-
5
System
Powered
No system problem
-
-
GREEN
6
System
Fault
Controller firmware non-functional
-
-
7
Channel 1
Fault
Damper not connected
-
GREEN
8
Channel 1
Fault
Channel shutdown due to 'Damper Fault'
-
GREEN
9
Channel 1
Fault
Channel shutdown due to 'Payload Change'
-
GREEN
10
Channel 1
Warning
Channel operating in 'overload' condition
-
GREEN
11
Channel 1
Tuning
Auto-tuning in process
B-GREEN
-
GREEN
12
Channel 1
Damping
Normal operation
GREEN
-
GREEN
13
Channel 1
Disabled
Channel shutdown via software
OFF
-
GREEN
14
Channel 2
Fault
Damper not connected
-
GREEN
15
Channel 2
Fault
Channel shutdown due to 'Damper Fault'
-
GREEN
16
Channel 2
Fault
Channel shutdown due to 'Payload Change'
-
GREEN
17
Channel 2
Warning
Channel operating in 'overload' condition
-
GREEN
18
Channel 2
Tuning
Auto-tuning in process
-
B-GREEN GREEN
19
Channel 2
Damping
Normal operation
-
GREEN
GREEN
20
Channel 2
Disabled
Channel shutdown via software
-
OFF
GREEN
AMBER
B-AMBER
AMBER
B-AMBER
AMBER
RED
RED
B-RED
RED
B-RED
Table 4
Definition of SYSTEM, CH1 and CH2 status LED’s.
4.1.3 I/O
The vibration can be monitored independently by connecting signal
processing equipment such as a signal analyzer or an oscilloscope to the BNC
jacks on the front panel of the controller. The measurement parameters are
listed in Table 5 below.
Measurement parameters
Sensitivity
30 V/(mm/s) ± 12%
Measurement range (0 – p)
0.16 mm/s
Frequency range (-3 dB)
18 Hz to 820 Hz
Table 5
Measurment Parameters