TP
read actual position
IMM
PGM
MIP
USAGE
♦
♦
SYNTAX
xx
TP
PARAMETERS
Description
xx
[ int ] -
axis number
Range
xx
-
1
to
MAX AXES
Units
xx
-
none
Defaults
xx
missing:
error 37, AXIS NUMBER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
DESCRIPTION
This command is used to read the actual position. It returns the instantaneous
real position of the specified axis.
RETURNS
nn
where:
nn
= actual position, in pre-defined units
REL. COMMANDS
PA
-
move to an absolute position
PR
-
move to a relative position
DP
-
read instantaneous desired position
EXAMPLE
3TP
|
read real position on axis # 3
5.322
| controller returns real position 5.322 for axis # 3
Section 3 – Remote Mode
3-141