Calibration, Maintenance, and Troubleshooting
1919-R User Manual
93
12.2.4
Updating Thermopile Sensor Calibration Factors
To update Thermopile sensor calibration factors:
1.
Press
Menu
and select
Calibration
. The screen displays the following general message:
Figure
12-2 General Calibration Information Screen
2.
Press
Continue
. The
Select Calibration Factor
screen displays the following calibration factors:
. Click the links for more details.
Figure
12-3 Select Calibration Factor Screen for Thermopile Sensors
3.
Select (highlight) a Calibration Factor (
Sensitivity, Energy
, or
Response
) and press
Continue
to
activate calibration based on the selected factor. The main Measurement screen appears,
displaying the relevant Factor values (on the left) based on the selected calibration factor. For
more details, see
, and
4.
Press
Modify
. The screen displays the
Original
(current) reading and factor (on the left side of
the screen), and
Updated
factor and its effect on the reading (on the right). The
Updated
box
on the right is highlighted, indicating it is modifiable. For more details, see
5.
In the
Updated
area, use the Right/Left arrows to select the digit to update. Use the Up/Down
arrows to change the value of the selected digit.
6.
Press
Save
to save the settings and return to the main Measurement screen, which now
displays the updated calibration factor values.
7.
Press
Exit
to return to the Menu screen.