558B
CURRENT LOOP INDICATOR
Operator’s Manual
http://www.newportUS.com/manuals
®
NEWPORT
Electronics, Inc.
Page 1: ...558B CURRENT LOOP INDICATOR Operator s Manual http www newportUS com manuals NEWPORT Electronics Inc...
Page 2: ...p www newportUS com is the policy of NEWPORT to comply with all worldwide safety and EMC EMI regulations that EWPORT is constantly pursuing certification of its products to the European New App irecti...
Page 3: ...M Factory Mutual 4 3 3 Installation Removal 5 4 0 SIGNAL INPUT CONNECTIONS P3 6 5 0 CONFIGURATION 7 5 1 Standard Setup 7 5 2 Field Configuration Chart 7 6 0 CUSTOMER CONFIGURATION AND CALIBRATION 9 6...
Page 4: ...ut the shipment use the phone number for the Customer Service Department nearest you When you receive the shipment inspect the container and equipment for signs of damage Note any evidence of rough ha...
Page 5: ...are used to display 1999 active counts plus one or two dummy right hand zeros The 558B displays from 1999 to 1999 counts with the option of one or two dummy right hand 0 digits Zero suppression or ele...
Page 6: ...center of each of the four overlapping zero ranges provided that the resultant is within the 1999 count display capability Span Slope Positive or Negative Normal Mode Rejection at 50 60 Hz 46 dB minim...
Page 7: ...rating 2 1 for each 10 C above 60 C Humidity derating 2 1 for each 10 RH above 60 RH 2 5 ENVIRONMENT Standard operating temp 0 to 55 C Extended operating temp option conformally coated 40 to 85 C Humi...
Page 8: ...e wires are securely clamped in the plug in connector by rotating the screws in the plug in connector clockwise but do not exceed a torque of 5 newton meter 37 pound foot Then plug the connector firml...
Page 9: ...e 3 2 Case Dimensions Insert the case in the panel cutout from the front and slide the sleeve on from the rear Install the two thumbnuts to secure the sleeve to the case Main Board Access 1 Case must...
Page 10: ...ows see Figure 3 1 P3 Connection Signal 1 Signal Hi 2 Signal Lo 3 No Connection 4 15 105 4 SLEEVE 3 50 88 9 1 89 48 0 3 78 96 0 7 9 0 31 MAX CASE TOP VIEW SIDE VIEW PANEL CUT OUT 45 00 0 61 0 00 R PAN...
Page 11: ...iguration option is specified jumpers are factory installed at S1 H S2 C S1 B S1 C and the unit is calibrated for 4 20 mA 00 0 to 100 0 5 2 FIELD CONFIGURATION CHART CONFIGURATION JUMPER INSTALLATION...
Page 12: ...8 Figure 5 1 Main Board Jumper Locations A A H J F C D E B G B J E H G R4 U3 U2 P3 VR1 S1 S2 S2 SPAN ADJ ZERO ADJ VR2 R30 S1 F C D...
Page 13: ...highest input current I2 which is specified by the customer I2 mA 3 Determine the reading N1 at input current I1 which is specified by the customer N1 counts 4 Determine the reading N2 at input curren...
Page 14: ...If a decimal point is required DP1 5 install a push on jumper as indicated in the configuration chart Section 5 2 6 If one dummy right hand zero is required install jumper at S2 A 7 If two dummy righ...
Page 15: ...eters apply Vmax 12 5V Imax 250mA Ci 8 6uF Li 0mH The Interconnecting Cable may be a twin pair or a pair contained in a type A or type B multicore cable provided that the peak voltage of any circuit c...
Page 16: ...rwise shall not exceed the purchase price of the component upon which liability is based In no event shall NEWPORT be liable for consequential incidental or special damages CONDITIONS Equipment sold b...
Page 17: ...88 www newportuk co uk e mail sales newportuk co uk Newport Electronics B V Postbus 8034 1180 LA Amstelveen The Netherlands TEL 31 20 3472121 FAX 31 20 6434643 Toll Free 0800 0993344 www newport nl e...
Page 18: ...mponents Click to view similar products for newport manufacturer Other Similar products are found below BDS1 BDT1 WTK 14 36 N 5TC TT J 24 72 TXDIN70 OTP U F 5SC TT K 30 36 WTK 8 24 N OTP U M TPJ U F 5...