Possible drawbacks found in the intraoral radiographs results
Clear image
Possible causes:
Exhausted development liquid
Excessive dilution of the developing liquid
Too short exposure time to X-rays
Insufficient development time
Exposure liquid temperature below the recommended range
Dark image
Possible causes:
Wrong dilution of developing liquid
Too long exposure time to X-rays
Excessive development time
Exposure liquid temperature higher than the recommended
Not detailed image
Possible causes:
Patient movement
Monoblock movement
Radiography partially exposed
Possible causes:
Error in the centering between rays’s beams and film
Liquid development too low with a consequent film partial development
Contact between two or more films during development
Veiled Image
Possible causes:
Films that exceeded the expiration date
Film accidental exposure to rays
Film accidental exposure to heat sources
Film accidental exposure to daylight or safety lamp darkroom no longer suitable