Rev. 00 of 12.06.09
15/18
Ca p.7
– MAIN TENA NCE
7 . 1
– K n o b
The connection cable must be periodically checked for cracks, possible cause
of dispersion.
7 . 2
– D e v i c e
Use a cloth moistened in water and alcohol to clean the device.
Do not use liquids
at all since there is not protection against their entrance (IP20)
.
7 . 3
– I m m e d i a t e m a i n t e n a n c e :
Immediate maintenance at New Age Italia or by authorised personnel if:
the device has been subject to external mechanical stress (e.g.. serious
drops);
the device has been overheated excessively (e.g. if left near intense heat
sources;
you doubt that liquids may have penetrated inside;
the power supply, the cover or other parts are damaged, broken or missing;
it appears that the features of the device have been changed.
for safety purposes, do not operate with accessories (e.g. knobs and
power supply) other than those included in the basic supply.
The frequency of maintenance, functional checkups and inspection for
compliance with safety standards EN60601-1 for medical devices carried out
with secur-tester is yearly. The useful life span of the instrument is
guaranteed by the company only if maintenance is carried out regularly.
N. B.:
have inspections carried out only by New Age Italia. The device can
be sent directly to company assistance workshops or delivered to the dealer
where the device was purchased.
Assistance centre:
New Age Italia srl -
Via De Brozzi, 3 - 48022 Lugo (RA)
Tel.:+39-0545.32019 - Telefax: +39-0545.369028
Web: