Rev. 02 del 24/07/13
10/14
Ca p.6
–
M AI NT EN AN CE
6 . 1
– H a n d l e
The handle and connection cable must be periodically checked for verifying that
there are no cracks; to clean the handle with a moist cloth and dry with a clean and
dry cloth.
6 . 2
– D e v i c e a n d p o w e r s u p p l y
To clean the device and the power supply cable we advises to use a moist cloth.
Don't use in any case liquids, because they are not protected from their entry (IP20).
6 . 3
– I m m e d i a t e m a i n t e n a n c e
We recommend you to perform the controls only from New Age Italia or from
the specialized personnel delegated if:
the instrument has been submitted to mechanical external solicitations , as
serious falls ;
the instrument has been submitted to strong overheating, for example, if left next
to intense heat sources;
you doubt that some liquids can be penetrated inside;
the wrap or other parts of the instrument are damaged, broken or lacking;
the functionality of the instrument appears altered.
To the goals of the safety we recommend you not to operate with accessories
(for example handle and power supply ) different from the ones furnished as
endowment of base.
The frequency of maintenance, of functional control and verification of
correspondence to the safety norms EN60601-1 for the medical devices, to perform
with secur-tester, is annual. The useful life of the tool is guaranteed only from the
firm if such maintenance is regularly effected.
ATTENTION:
We recommend you to perform the controls only from New Age
Italia or from the specialized personnel delegated by this Firm. The instrument in
maintenance can be sent directly to the assistance laboratories or delivered to the
reseller by which it has been purchased.