Programming Instructions
75
RNS
®
System User Manual
S
ETTING UP
R
ESPONSIVE
T
HERAPY
The RNS
®
Neurostimulator can deliver a therapy sequence of up to 5 individually configured
therapies in response to detected activity. The neurostimulator resumes sensing after each therapy is
delivered. If the activity is still detected, the next (sequential) therapy is delivered. If the activity is no
longer detected, the remaining therapies are not delivered and the episode ends.
Note: The neurostimulator will deliver up to 5 sequential therapies within an episode. Once 5
therapies have been delivered, the neurostimulator will wait for the episode to end and detect
new activity before additional therapies can be delivered.
Note: It is recommended to review and perform detection analysis prior to enabling responsive
therapy.
Behavior for the first therapy in the therapy sequence may be configured in one of two ways:
•
Different bursts may be delivered in response to a specific event detector (Pattern A or
Pattern B) as depicted in Figure 34; or
•
Up to two bursts may be selected to be delivered in response to any event detector as
depicted in Figure 35.
All subsequent therapies deliver a maximum of 2 bursts and are independent of the event detector.
Figure 34:
A
T
HERAPY
S
EQUENCE
D
ELIVERED
I
N
R
ESPONSE
T
O
A
S
PECIFIC
E
VENT
D
ETECTOR
Therapy 1 is configured to deliver a single burst in response to a specific detector. In this example, Pattern A
is detected at the episode start; Therapy Pattern A burst is delivered. If Pattern B is detected at the episode
start, Therapy Pattern B burst is delivered. Subsequent therapies, Therapy 2 and Therapy 3, are configured to
deliver 2 bursts each, no matter which pattern is subsequently detected.
Figure 35:
A
T
HERAPY
S
EQUENCE
D
ELIVERED
I
N
R
ESPONSE
T
O
A
NY
E
VENT
D
ETECTOR
Therapy 1 is configured to deliver Burst 1 and Burst 2 in response to any event detector. In subsequent
therapies, Therapy 2 is configured to deliver Burst 1 only and Therapy 3 is configured to deliver Burst 1 and
Burst 2.