Chapter 5 - Software Reference Manual
43
Figure 24: Display during stimulation
Terminating the stimulation
The ongoing stimulation can be terminated by the user at any time by pressing button 1 (x).
The display indicates: “Stimulation terminated by user!”. The menu then returns to the main
menu PARAMETER.
Impedance control
The device permanently measures the electrode impedance during stimulation. If the
predefined impedance limit is exceeded during stimulation an error warning occurs and the
stimulation will pause immediately. Press button 1 to go back to the main menu
PARAMETER.
Completing the stimulation
The stimulation stops automatically after the predefined time is over. The menu returns to
the main menu PARAMETER.
SYSTEM
In the Software Reference Manual the chapter Trigger input only describe
software settings. For technical details please see Manual: Operation Basics.