Sphygmomanometer Calibration Verification Guideline 

 

200-BP-CAL-KIT

Sphygmomanometer

 Calibration Kit 

INSTRUCTION MANUAL

公司:深圳市净康科技有限公司

地址:深圳市龙岗区沙平北路111号6008

 

联系人:王小姐      电话

: 15813841944  0755-28917660  

QQ

:422612157   邮箱:[email protected]

www.300buy.com

 

Summary of Contents for 200-BP-CAL-KIT

Page 1: ...libration Verification Guideline 200 BP CAL KIT Sphygmomanometer Calibration Kit INSTRUCTION MANUAL 公司 深圳市净康科技有限公司 地址 深圳市龙岗区沙平北路111号6008 联系人 王小姐 电话 15813841944 0755 28917660 QQ 422612157 邮箱 jkang66 163 com www 300buy com ...

Page 2: ...ion replace at no charge a product that proves to be defective provided you return the product shipping prepaid to Netech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as the result of service or modification by other than Netech Corporation or if its serial number is defaced or removed Netech reserves the right to discontinue the Blood Pressure ...

Page 3: ... 1 General Overview Introduction 05 Specifications 05 Accessories 06 Optional Accessories 06 2 Operating Instructions Controls and Indicators 07 Typical Calibration Setup 08 3 Troubleshooting The Calibration Setup Maintaining Initial Zero 09 Maintaining Steady Pressures 10 ...

Page 4: ...nits and pressure ranges are available To provide superior performance the Digimano 1000 incorporates one of the latest semiconductor pressure sensors It is powered by one 9 Volt alkaline battery Optionally an AC adapter may be used The Digimano 1000 is CE marked and is shipped with a Certificate of Calibration traceable to the NIST The Digimano 1000 is a rugged instrument that performs its measur...

Page 5: ...O 2000IN 1999 mmHg 199 9 inH2O 200PS 199 9 mmHg 19 99 PSI 2000PS 1999 mmHg 19 99 PSI 200cmH2O 199 9 cmH2O 19 99 PSI 200kPa 199 9 kPa 1999 cmH2O 2000kPa 199 9 kPa 199 9 inH2O User can pick any of these models for the BP Cal Kit 1 3 ACCESSORIES Description Part Number User Manual 200 Manual Syringe Pressure Applicator 204 Silicon Tubing Three way Plastic TEE Connector Hard Carrying Case 1 4 OPTIONAL...

Page 6: ...CONTROLS AND INDICATORS 1 Display Screen 2 ON OFF Switch 3 mmHG PSI Switch 4 Luer Lock Fitting Female 5 DC Proportional Output 6 Zero Adjustment 7 Lead Screw Release 8 Precision Adjust 9 Luer Lock Fitting Female 10 Barb Fitting 11 Zero Square Digimano Pressure Meter Figure 1 3 5 4 2 1 6 ...

Page 7: ...indicator 11 see Figure 2 If not the unit requires repair prior to proceeding Turn your digital pressure meter on and insure that the display 1 reads zero if it does not adjust to zero by using the zero adjustment See Figure 1 Adjust the plunger on the Pressure Applicator Syringe to to its mid point by pushing or pulling the plunger 8 while pressing in the lead screw release button 7 see Figure 2 ...

Page 8: ... 3 3 0 TROUBLESHOOTING THE CALIBRATION SET UP 3 1 Maintaining Initial Zero readings The sphygmomanometer and or the digital pressure meter reads a pressure after connecting all the tubing and devices Make sure that you use the male luer lock fitting when connecting either the digital pressure meter or the pressure applicator syringe Always connect the sphygmomanometer first and the syringe last us...

Page 9: ...ubing segment c at the syringe If the pressure remains steady tighten up the syringe connection If this fails clamp tube segment c at the TEE Connector if pressure stays steady tighten up this connection Else clamp segment b at the TEE connector If pressure stays steady tighten up this connection Else clamp segment b at the Digital Pressure Meter connection If pressure stays steady tighten up this...