WP-Series Reverse Osmosis System
Installation Instructions and User Manual
Page | - 10 -
System Installation
Location
Locate R.O. system with adequate clearance from walls and other equipment to enable membrane servicing. Run three
polyvinyl tube lines to the system as follows:
•
Line #1
: Connect raw water feed supply to the solenoid valve inlet in front of the first cartridge filter housing. This
will require ½ inch threaded pipe fitting and adaptor for the feed line. If desired, install an isolation valve in this
line; ensure the valve opening does not restrict the water flow.
Illustration 3 - Line 1 Connection
CAUTION:
Pump and system performance will be adversely affected if the feed/suction line is
restricted.
•
Line #2:
Run a 3/8" line from the open end of the concentrate valve to a drain. Ensure that no liquids from other
lines in the plant flow back through this line.
Illustration 4 - Line 2 & 3 Connections
•
Line #3:
Run a 3/8" line from the permeate outlet to desired permeate storage container. Pressurized storage or
atmospheric storage tank options are available separately.
Note: The recirc flow line is pre-connected at the factory.
CAUTION:
Be sure that all lines are connected before plugging in unit power.
Summary of Contents for WP Series
Page 31: ...Notes ...