Installation Guide
Part #:
70-0285
(See DSU-H Instructions for Use)
Questions or issues?
Let us help!
Call:
(201) 343-5202, Ext. 1
MACHINE
DISCONNECT
DISCONNECT
OFF
MACHINE
CONNECT
1/2” NPS or NPT
CONNECT
1
2
3
READ the entire installation guide and
precautions before starting.
WASH hands before starting installation.
WEAR gloves before handling filter (do not
remove blister pack).
TURN OFF the water supply valve.
DISCONNECT the supply line from the
machine’s inlet and water valve.
CONNECT a new flex line to the machine’s inlet.
CONNECT the quick connector with bushing to
the other end of flex line.
MACHINE
CONNECT
INLET
BLEACH
PULL
PUSH
CONNECT
CONNECT
1/2” NPS or NPT
MALE THREAD
4
5
6
Clorox
®
Healthcare Bleach Germicidal Cleaner is
the recomended disinfectant choice.
CONNECT the outlet of machine disinfection unit
to the machine flex line quick connector.
PLACE the inlet hose of unit into disinfectant
(e.g. BLEACH) container.
PLACE a cup under the water dispenser and
keep OPEN the dispenser. This is to vent the
machine and allow for disinfectant to fill.
PULL out syringe plunger to draw the
disinfectant.
PUSH the plunger to deliver the disinfectant and
continue until disinfectant is seen coming out of
the machine’s dispenser.
ALLOW disinfectant to dwell for 5 minutes.
CONNECT a new flex line to the water valve.
CONNECT the quick connector with bushing to
the other end of flex line.
TURN ON the water source to assure there are
no leaks, then turn the water source back off.
DSU-H In-Line
AVOID contact between the filter and cleaning or sanitizing agents.
(1)
DO NOT remove blister pack from filter until indicated to protect from
contamination.
MACHINE
MACHINE
DISCONNECT
CONNECTION
CONNECTION
CONNECT
CONNECT
DISCONNECT
OFF
INLET
MACHINE
CONNECT
DISINFECTANT
PUSH
PULL
(1) Medical grade plastics are more susceptible to cracking due to exposure to organic solvents.
Alkyl dimethyl benzyl ammonium chloride (ADBAC), glycols, alcohols, ketones, silicones, and
acetones are some examples of the leading organic solvents found in cleaning and sanitizing
agents. Use of cleaning/sanitizing agents on the filter housings is not required and should be
avoided.