Dry Bath N10 Series Operation manual Chapter 5 Failure Analysis and Troubleshooting
—
13
—
Chapter 5 Failure Analysis and Troubleshooting
Problems and actions
No.
Common problem
Possible cause
Action(s)
1
No
display
on
the
screen
No power on the main
power plug
Check power supply and
plugged properly
Faulty fuse
Change fuse
On/Off button broken
Change button
Others
Contact us
2
The
actual
and
displayed temperatures
are quite different
Broken
sensor
or
loose contact of the
block
Contact us
3
“OPEn” displayed,
alarming “beep”
Sensor disconnect
Contact us
“SHOr” displayed,
alarming “beep”
Sensor short-circuit
“HHHH”
displayed,
alar
ming “beep”
Sensor broken, or
block temperature is
too high
4
No heating
Sensor broken
Contact us
Heating tube broken
5
Keys don’t work
Faulty key
Contact us