23
3.
The device heats up:
The continuous operation
should be at most 25 minutes long. Take a 40min
break before using the device again.
4.
The appliance operates too loudly:
The air tube is incorrectly installed.
The air filter cover is not properly secured.
11.
ELECTROMAGNETIC COMPATIBILITY
1.
This appliance must be installed and put into service in accordance with the
information contained in the following instructions.
2.
The range of tests for the basic safety and essential performance of medical
equipment and systems should be selected based on a high probability of
maintaining basic safety and essential performance and should be com
-
patible with the professional healthcare facility environment, the home
healthcare environment and the specialty environment, based on the
locations of intended use.
3.
Home healthcare is the residence where the patient lives or other places
where patients reside, excluding professional healthcare settings where
medically trained operators are permanently available when patients are
present. Such as schools, outdoor facilities, homes, hotels.
4.
Please dispose of parts of the appliance at the end of its use in accordance
with the rules in force at the location.
Deduction and generation of electromagnetic emissions
The NENO SANO Compressor Nebuliser (VP-D2) is designed for use in the elec
-
tromagnetic environment described below; the customer or user of the nebuliser
should ensure that it is used in such an environment.
Emissions
testing
Compatibility
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The compressor nebuliser NENO SANO
(VP-D2) uses RF energy only for its internal
functions. Therefore, RF emissions are very
low and do not cause interference to nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The NENO SANO (VP-D2) compressor nebu
-
liser is suitable for use in all non-domestic
premises and those directly connected to the
public low-voltage mains supplying buildings
used for residential purposes.
Harmonic
emission
IEC 61000-3-2
Class A
Summary of Contents for Sano
Page 1: ...Sano ...
Page 2: ......
Page 3: ...3 A B 1 1 8 8 4 4 2 2 9 5 5 3 3 10 6 6 7 7 11 ...
Page 4: ...4 1 6 3 2 5 10 4 14 12 9 8 C 13 11 7 1 2 D 3 4 5 ...
Page 5: ...5 6 7 1 2 E ...
Page 6: ...6 4 1 2 3 F ...
Page 116: ...PL DE CZ EN ...
Page 117: ...HU LT SK LV ...
Page 118: ...RO ET BG ...
Page 119: ......
Page 120: ......