Table 11: Electromagnetic Immunity
The NPB-40 is intended for use in the electromagnetic environment
specified below. The customer or user of the NPB-40 should assure that
it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
Complies
Floors should be
wood, concrete, or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
+ 2 kV power
supply lines
+ 1 kV for
input/output
lines
Complies
Main power should
be that of a typical
commercial or
hospital
environment.
Power
Frequency
(50/60 Hz)
magnetic field
3 A/m
Complies
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Summary of Contents for Oximax NPB-40
Page 1: ......
Page 8: ...Blank Page...
Page 14: ...Blank Page...
Page 32: ...Blank Page...
Page 38: ...Blank Page...
Page 80: ...Blank Page...
Page 82: ...NPB 40 Operation 74...
Page 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Page 104: ...Blank Page...
Page 105: ...Performance Considerations NPB 40 97...
Page 106: ...Performance Considerations 98...
Page 116: ...Blank Page...
Page 120: ...Blank Page...
Page 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Page 126: ...Blank Page...
Page 132: ...Blank Page...
Page 151: ...Specifications NPB 40 143...
Page 152: ...Specifications 144...
Page 159: ...Index NPB 40 153...
Page 160: ......