Technical Information
86
Factory Defaults
The N
-
560 is shipped with factory default settings. Authorized technical
personnel using the procedures described in the N
-
560 service manual can
change default settings.
Table 5: Factory Default Settings
Parameter
Range
Factory Default
Setting
%SpO
2
Upper Alarm
Limit
Lower Alarm Limit plus 1
to 100%
100%
%SpO
2
Lower Alarm
Limit
20% to Upper Alarm
Limit minus 1
85%
Pulse Rate Upper Alarm
Limit
Lower Alarm Limit plus 1
to 250 bpm
170 bpm
Pulse Rate Lower Alarm
Limit
30 bpm to Upper Alarm
Limit minus 1
40 bpm
Alarm Silence Reminder
On or Off
On
Alarm Silence Duration
Off, 30, 60, 90, 120
seconds
60 seconds
Alarm Silence
Restrictions
• Audible reminder
• No audible reminder
• Do not allow alarms off
Audible reminder
Alarm Sound Selector
1, 2, 3
1
Alarm Volume
1 to 10
4
Data Port Baud Rate
2400, 9600, 19200
19200
Data Port Protocol
1, 2
1 (ASCII)
Display
0, 1
0 (On)
In-Sensor Trend Mode
0, 1, 2
0 (Event SpO
2
)
Language
English
English
Pulse Beep Volume
0 to 10
4
RS-232 Nurse Call
Priority
Normally high, normally
low
Normally low
SatSeconds
Off, 10, 25, 50, 100
Off
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......