Te c h n i c a l I n f o r m a t i o n
Description of Alarms
The N
-
560 has three levels of audible alarms.
1.
High
-
priority alarm
: A high
-
pitched, fast
-
pulsating tone indicating
loss of pulse with no patient motion.
2.
Medium
-
priority alarm
: A medium
-
pitched, normal
-
pulsating tone
indicating an SpO
2
or pulse rate limit violation.
During a medium
-
priority alarm, the green display turns red and
flashes with the patient parameter that violated the limit (%SpO
2
or
Pulse Rate). If the alarm is a
SatSeconds
alarm, the
SatSeconds
indicator (clock) will flash.
3.
Low
-
priority alarm
: A low
-
pitched, slow
-
pulsating tone indicating a
sensor disconnect, low battery, or N-560 failure.
During a low
-
battery alarm, the
Low Battery
indicator illuminates
and the alarm tone sounds immediately, even if the alarms are
silenced or set to OFF.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......