Performance Considerations
N-560 Operator’s Manual
69
P e r f o r m a n c e C o n s i d e r a t i o n s
WARNING: Pulse oximetry readings and pulse signals can be
affected by certain ambient environmental conditions, sensor
application errors, and certain patient conditions. See the
appropriate sections of the manual for specific safety information.
Performance Verification
The performance of the N
-
560 can be verified by following the
procedures outlined in the Performance Verification section of the N
-
560
service manual. Qualified service personnel should perform these
procedures before using the N-560 for the first time in a clinical setting.
N-560 Performance Considerations
Certain patient conditions can affect the measurements of the N
-
560 and
cause the loss of the pulse signal.
Inaccurate measurements can be caused by:
•
excessive patient movement
•
venous pulsations
•
intravascular dyes, such as indocyanine green or methylene blue
•
defibrillation
Dysfunctional Hemoglobins
Dysfunctional hemoglobins such as carboxyhemoglobin, methemoglobin,
and sulphemoglobin are unable to carry oxygen. SpO
2
readings may
appear normal; however, a patient may be hypoxic because less
hemoglobin is available to carry oxygen. Further assessment beyond
pulse oximetry is recommended.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......