![Nellcor OXIMAX N-560 Operator'S Manual Download Page 108](http://html1.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_operators-manual_3908427108.webp)
Specifications
100
Environmental Conditions
Operation
Temperature
50 ºF to 113 ºF (10 ºC to 45 ºC)
Altitude/Barometric Pressure
-
390 m to 3,012 m
(
-
1,280 ft. to 9,882 ft.)
70 kPa to 106 kPa
(20.6 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-
condensing to be
compliant with IEC 60601
-
1, sub
-
clause
44.5
Transport and Storage (not in shipping container)
Temperature
-
4 ºF to 140 ºF (
-
20 ºC to 60 ºC)
Altitude/Barometric Pressure
-
390 m to 5,574 m
(
-
1,280 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-
condensing over
temperature range of
-
4 ºF to 140 ºF
(
-
20 ºC to 60 ºC)
Transport and Storage (in shipping container)
Temperature
-
4 ºF to 158 ºF (
-
20 ºC to 70 ºC)
Altitude/Barometric Pressure
-
390 m to 5,574 m
(
-
1,280 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-
condensing
Sensor Power Dissipation
Sensor
Dissipation
O
XI
M
AX
MAX-N
52.5 mW
O
XI
M
AX
MAX-I
52.5 mW
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......