Specifications
N-560 Operator’s Manual
95
S p e c i f i c a t i o n s
Performance
Measurement Range
SpO
2
1% to 100%
Pulse Rate
0 and 20 beats per minute (bpm) to 250 bpm
Perfusion Range
0.03% to 20%
Accuracy and Interference Tolerance
Saturation
Adults
1
70 to 100% ±2 digits
Neonate
70 to 100% ±3 digits
Low Perfusion
2
70 to 100% ±2 digits
Pulse Rate
Adult and Neonate
1
20 to 250 bpm ±3 digits
Low Perfusion
2
20 to 250 bpm ±3 digits
1
Adult specifications are shown for OXIMAX MAX-A and MAX-N
sensors with the N-560. Neonate specifications are shown for
OXIMAX MAX-N sensors with the N-560. Saturation accuracy will
vary by sensor type. Refer to the Sensor Accuracy Grid. The Sensory
Accuracy Grid is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/Product-
Manuals.html
2
Specification applies to N-560 performance. Reading accuracy in the
presence of low perfusion (detected IR pulse modulation
amplitude <1.5%) was validated using signals supplied by a
patient simulator. SpO
2
and pulse rate values were varied across the
monitoring range including weak signal conditions and compared to
the known true saturation and pulse rate of the input signals.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......