Specification
N5500 Service Manual 77
Technical
Information, N5500
SpO
2
%Saturation
Range
1% to 100%
Low Perfusion
0.03% to 20%
Accuracy
Without Motion-Adults
1
70% to 100% ±2 digits
Without Motion-Neonate
1
70% to 100% ±3 digits
1% to 69% unspecified
With Motion
2
-Adults/Neonate 70% to 100% ±3 digits
Low Perfusion
3
70% to 100% ±2 digits
1% to 69% unspecified
Display Sweep Speeds
12.5 mm/sec, 25.0 mm/sec, and 50.0 mm/sec
Pulse Rate
Range
20 BPM to 250 BPM
Accuracy
Without Motion
3
20 BPM to 250 BPM ±3 digits
With Motion normal physiologic range
(55 to 125 BPM) ±5 digits
Low Perfusion
3
20 BPM to 250 BPM ±3 digits
1
Adult specification are shown for OxiMax MAX-A and MAX-N sensors with the N5500. Neonate
specifications are shown for OxiMax MAX-N sensors with the N5500. Saturation accuracy will vary
by sensor type.
2
Applicability(accuracy with motion): OxiMax sensors.
3
Specification applies to monitor performance and was validated with Biotek and Nellcor simulators
Temperature
Probe Type
Thermistor probe (Monotherm™ temperature probes)
Parameter displayed
T1, T2
Range
15° C to 45° C
Display Accuracy
±0.1° C (25° C to 45° C)
±0.2° C (15° C to less than 25° C)
Probe Accuracy
±0.1° C
9.8 Trends
Types
Graphical and Tabular
Memory Storage
24 hours, nonvolatile, 20 seconds data interval
Graphical Format
One graph for all parameters
Also user-selectable each parameter to be desired
Display Range
2 hours
Tabular Format
One table for all variables.
18 rows (date, parameter labels, and 16 variable rows)
Display Time Interval 20 sec, 1 min, 2 min, 3 min, 5 min, 10 min, 15 min, 20 min