N-550
81
S p e c i f i c a t i o n s
Performance
Measurement Range
SpO2
1% to 100%
Pulse Rate
0 and 20 beats per minute (bpm) to
250 bpm
Perfusion Range
0.03% to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion
-
Adults
1
70 to 100% ±2 digits
Without Motion
-
Neonate
1
70 to 100% ±3 digits
With Motion
-
Adults and Neonates
2
70 to 100% ±3 digits
Low Perfusion
3
70 to 100% ±2 digits
Pulse Rate
Without Motion
3
20 to 250 bpm ±3 digits
With Motion
normal physiologic range (55
-
125 bpm) ±5 digits
Low Perfusion
3
20 to 250 bpm ±3 digits
1 Adult specifications are shown for
O
XI
M
AX
MAX
-
A and MAX
-
N sensors with
the N
-
550. Neonate specifications are shown for
O
XI
M
AX
MAX
-
N sensors with
the N
-
550. Saturation accuracy will vary by sensor type. Refer to the Sensor
Accuracy Grid.
2
Applicability:
O
XI
M
AX
MAX
-
A, MAX
-
AL, MAX
-
P, MAX
-
I, and MAX
-
N
sensors.
3
Specification applies to N-550 performance.
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...