Performance Verification
N-550
27
on page 22. The pulse amplitude indicator should
indicate low level modulation.
on page 23. The pulse amplitude indicator should
indicate low level modulation.
11. Disconnect all equipment and turn off the N-550.
Safety Tests
The N-550 safety tests meet the standards of, and are performed in accordance
with, IEC 60601-1 (EN 60601-1, Amendment 1, Amendment 2) and UL
2601-1, for instruments classified as Class 1 and TYPE BF and ANSI/AAMI
Standard ES1.
Applicable tests for these standards are listed below. Technicians must be
familiar with the standards applicable to their institution and country. Test
equipment and its application must comply with the applicable standard.
•
Ground Integrity, see page 87 for test values.
•
Earth Leakage Current, see page 87 for test values.
•
Enclosure Leakage Current, see page 87 for test values.
•
Patient Applied Risk Current, see page 88 for test values.
•
Patient Isolation Risk Current, see page 88 for test values.
Note
:
Patient Applied Risk Current and Patient Isolation Risk Current. The
leakage test lead from the test equipment must be connected to the
N-550 SpO
2
Sensor Port using a male 9-pin “D” type connector that
has all pins shorted together.
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...