Performance Verification
20
Introduction
The SRC-MAX functional tester allows qualified technicians to functionally
test Nellcor O
XI
M
AX
technology-based pulse oximeters and OEM O
XI
M
AX
technology-based monitors. The technician must perform the test setup
procedure before performing tests 1 through 4. The following is a brief
description of each test:
•
Test Setup
— This procedure establishes the baseline for all the other
tests. The Test Setup procedure must be performed before performing
any or all of the SRC-MAX tests.
•
Test 1: BPM
— This procedure simulates an O
XI
M
AX
sensor
attached to a patient indicating 60 bpm and 200 bpm. The test setup
procedure sets up Test 1 for 60 bpm.
•
Test 2: SpO
2
— This procedure simulates an O
XI
M
AX
sensor
attached to a patient, indicating 75 percent blood oxygen saturation
and 90 percent blood oxygen saturation. The test setup procedure sets
up Test 2 for 75 percent blood oxygen saturation.
1 — DOC-10 Cable Connector
6 — % Modulation Select Button
2 — Red LED Drive Indicator
7 — % SpO
2
Select Button
3 — Not Used For N-550
8 — Light Level Selection Button
4 — Not Used For N-550
9 — Pulse Rate Selection Button
5 — Battery Low Indicator
10 — Infrared LED Drive Indicator
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...